ZOLL IVTM THERMOGARD XP 8700-0650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-05 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.

Event Text Entries

[186584966] An onsite evaluation of the thermogard xp ivtm console (sn (b)(4)) was performed by a zoll service technician. The report of the thermogard console was slow to cool was confirmed during functional testing. The root cause is due to a dirty air filter and dirty tank due to lack of preventive maintenance. Upon visual inspection, tank and air filter was dirty. Unrelated to the reported complaint, the console's coldwell cap assembly and drip pan was missing. A review of the event log was performed and no issues were observed. After service and replacing air filter, coldwell cap assembly and drip pan, the console passed all testing criteria and meets performance specifications. Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for the thermogard console with serial number (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186584968] The thermogard console (sn (b)(4)) was slow to cool. The reporter was unable to specify if the issue occurred during device check or patient use. No consequences or impact to patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00228
MDR Report Key9794345
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-05
Date of Report2020-03-05
Date of Event2020-02-07
Date Mfgr Received2020-02-10
Device Manufacturer Date2016-07-07
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM THERMOGARD XP
Generic NameTHERMAL REGULATING SYSTEM
Product CodeNCX
Date Received2020-03-05
Model Number8700-0650
Catalog Number8700-0650
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.