XPS? CLEARVIEW - LEGEND TOOL SP14MH30T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-05 for XPS? CLEARVIEW - LEGEND TOOL SP14MH30T manufactured by Medtronic Xomed Inc..

Event Text Entries

[182211806] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[182211807] A health care provider (hcp) reported that during navigated transforaminal lumbar interbody fusion the burr seemed to not be advancing through the pedicle having been used to make 2 holes already with no issues. When the surgeon looked at the end of the burr, it was noticed the tip of the burr was missing. An additional intraoperative computerized tomography was conducted in order to find th e broken piece of burr. The surgeon decided that as the piece of burr was within the pedicle and not likely to become loose, to leave the piece in place as it would cause more damage to remove the broken piece. The procedure was completed and at the end of the procedure another ct spin was completed and the broken end was still within the middle of the pedicle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00138
MDR Report Key9794407
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-05
Date of Report2020-03-27
Date of Event2020-02-13
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-11-12
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXPS? CLEARVIEW - LEGEND TOOL
Generic NameDRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Product CodeHBE
Date Received2020-03-05
Model NumberSP14MH30T
Catalog NumberSP14MH30T
Lot Number0218914695
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2020-03-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.