MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-05 for PENUMBRA SYSTEM JET7 REPERFUSION CATHETER 5MAXJET7KIT-B 5MAXJET7KIT manufactured by Penumbra, Inc..
[188508137]
The device has been returned and the investigation results are pending. A follow up mdr will be submitted upon completion of the device investigation.
Patient Sequence No: 1, Text Type: N, H10
[188508138]
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), neuron max 6f 088 long sheath (neuron max), and non-penumbra stent retriever. During the procedure, the physician advanced the stent retriever and jet7 to the face of the clot in the target vessel using the? Grappling hook? Technique. To complete the first pass, the stent retriever was then retracted, followed by the jet7. While retracting the stent retriever and jet7 through the rotating hemostasis valve (rhv), they became stuck. It was noted that the rhv was open all the way. Subsequently, it was entirely removed, and the physician was able to pull the stent retriever and jet7 out. Upon visual inspection, the jet7 was observed to be stretched and bent about 2 cm from the distal end. The procedure was completed using another non-penumbra stent retriever, non-penumbra catheter, same sheath, and balloon angioplasty. There was no report of an adverse effect to the patient.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005168196-2020-00309 |
MDR Report Key | 9795636 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-05 |
Date of Report | 2020-02-12 |
Date of Event | 2020-02-12 |
Report Date | 2005-01-01 |
Date Reported to FDA | 2005-01-01 |
Date Reported to Mfgr | 2005-01-10 |
Date Mfgr Received | 2020-02-12 |
Date Added to Maude | 2020-03-05 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. VERONICA FARRIS |
Manufacturer Street | ONE PENUMBRA PLACE |
Manufacturer City | ALAMEDA CA 94502 |
Manufacturer Country | US |
Manufacturer Postal | 94502 |
Manufacturer Phone | 5107483200 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PENUMBRA SYSTEM JET7 REPERFUSION CATHETER |
Generic Name | DXE |
Product Code | DXE |
Date Received | 2020-03-05 |
Returned To Mfg | 2020-02-28 |
Model Number | 5MAXJET7KIT-B |
Catalog Number | 5MAXJET7KIT |
Lot Number | F94085 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PENUMBRA, INC. |
Manufacturer Address | ONE PENUMBRA PLACE ALAMEDA CA 94502 US 94502 |
Product Code | --- |
Date Received | 2020-03-05 |
Device Sequence No | 101 |
Device Event Key | 0 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-05 |