INTRALASE PI-RET

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-05 for INTRALASE PI-RET manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[182566025] The patient interface (pi) suction ring may lose suction during a procedure. Label copy states corneal fixation vacuum loss can occur. There are several factors that may contribute to suction issues such as doctor? S technique in applying the suction ring to the cornea, doctor? S technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient? S cornea and the suction ring. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[182566026] During the raster pattern we lost suction right before the side cut. Doctor released pedal. Surgeon followed the suction break protocol. Re-applanated took pocket off. Centered the laser over raster pattern and completed flap. While the doctor was lifting he was having trouble getting the flap up and looked like he might be on the wrong plane. He decided to stop lifting the flap.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006695864-2020-00161
MDR Report Key9796150
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-05
Date of Report2020-03-05
Date of Event2019-12-19
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-10
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Street510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameINTRALASE
Generic NameFEMTOSECOND LASER
Product CodeHNO
Date Received2020-03-05
Model NumberPI-RET
Catalog NumberPI-RET
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E ST ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-05
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-05

© 2024 FDA.report