MINIMAL ACCESS ATTACHMENT MA-D20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-06 for MINIMAL ACCESS ATTACHMENT MA-D20 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[182405422] (b)(4). Device evaluation: the actual device was returned for evaluation. During repair, it was determined that the reported condition was confirmed. The assignable root cause was determined to be due to component failure from normal wear. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: N, H10


[182405423] It was reported by (b)(6) that during service and evaluation, it was determined that the bearing of the attachment device was damaged. It was observed that a spare part of the device was damaged during disassembly, and the ball bearing was defective. It was further determined that the device failed pretest for max temperature of attachment. It was noted in the service order that the device did not work during pre-surgery. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown. However, it was reported that the event occurred in 2020. All available information has been disclosed. If additional information should available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045834-2020-00397
MDR Report Key9798960
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-06
Date of Report2020-01-09
Date of Event2020-01-01
Date Mfgr Received2020-02-20
Device Manufacturer Date2015-09-15
Date Added to Maude2020-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal33410
Manufacturer Phone6103142063
Manufacturer G1DEPUY SYNTHES PRODUCTS LLC
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal Code33410
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMINIMAL ACCESS ATTACHMENT
Generic NameDRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Product CodeHBE
Date Received2020-03-06
Returned To Mfg2020-01-30
Model NumberMA-D20
Catalog NumberMA-D20
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer Address4500 RIVERSIDE DRIVE PALM BEACH GARDENS FL 33410 US 33410


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-06

© 2026 FDA.report