OLYMPUS A5294A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05,06 report with the FDA on 2008-01-16 for OLYMPUS A5294A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[745984] The user facility reported that after the procedure, it was noted that the patient sustained two second-degree burns under the patient's navel. The user facility claimed that the source of the burns came from the light cord used with the telescope, but the user also claimed that the heat originated from the telescope and not the light cord. The user also reported that the light cord was tested after the procedure and there was no problem found. A regular burn dressing was applied to the affected area and the patent was released from the facility the same day. The patient was reportedly doing fine.
Patient Sequence No: 1, Text Type: D, B5


[7930016] The device referenced in this report was returned to olympus for investigation. The investigation confirmed that the light guide post of the telescope became hot after testing the telescope for an hour with the use of an olympus test light source and an olympus test light guide cable. The light guide post temperature was measured twice and the temperature readings were 44 & 38 degree celsius. In addition, the telescope had nicks and dents, and the outer tube was found sharp due to deep scratches. This phenomenon was isolated to physical damage and user handling. This report is being filed as an mdr in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2008-00004
MDR Report Key979913
Report Source00,05,06
Date Received2008-01-16
Date of Report2007-12-18
Date Added to Maude2008-10-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameTELESCOPE
Product CodeFBP
Date Received2008-01-16
Model NumberA5294A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key983791
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressHAMBURG


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-16

© 2024 FDA.report