MYSPINE MYSPINE LP DRILL PILOT GUIDE L04 7.0554

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-06 for MYSPINE MYSPINE LP DRILL PILOT GUIDE L04 7.0554 manufactured by Medacta International Sa.

Event Text Entries

[188933600] Batch review performed on 18 february 2020: lot 02581s: (b)(4) item manufactured and released on 23-jan-2020. No anomalies found related to the problem. Patient match planing review each step of the process have been analyzed by my solution department. We analyzed each step of the myspine process. Images qc: the images we received were subjected to quality control, the result of which was positive. Reconstruction: the reconstruction is precise and correct, all areas were clearly identifiable. Planning: the surgeon modified and validated the planning (rev. 1). Guide: the guide was designed according to the trajectory of the screws in the planning validated (rev. 1). As usual, during the guide design we checked with the upper and the lower vertebra that there was no impingement. Conclusions our analysis of the myspine process of this case found no deviations from the standard procedures. Each step has been performed correctly
Patient Sequence No: 1, Text Type: N, H10


[188933601] It was not possible to use myspine guide l4 because there was a problem related to its settng. The patient lamina of l4/l5 were too closer, so in order to set the myspine guide l4, the patient lamina l4 should have been recected. The surgery finished without using myspine guide. 30 minutes delay and additional anesthesia has been used.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005180920-2020-00131
MDR Report Key9799371
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-06
Date of Report2020-02-06
Date of Event2020-02-06
Date Mfgr Received2020-02-06
Device Manufacturer Date2020-01-23
Date Added to Maude2020-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR STEFANO BAJ
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, SWITZERLAND 6874
Manufacturer CountrySZ
Manufacturer Postal6874
Manufacturer G1MEDACTA INTERNATIONAL SA
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, 6874
Manufacturer CountrySZ
Manufacturer Postal Code6874
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMYSPINE MYSPINE LP DRILL PILOT GUIDE L04
Generic NameDRILL PILOT GUIDE
Product CodeMNI
Date Received2020-03-06
Catalog Number7.0554
Lot Number02581S
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDACTA INTERNATIONAL SA
Manufacturer AddressSTRADA REGINA CASTEL SAN PIETRO, 6874 SZ 6874


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.