AQUABEAM ROBOTIC SYSTEM AB2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-06 for AQUABEAM ROBOTIC SYSTEM AB2000 manufactured by Procept Biorobotics Corporation.

Event Text Entries

[188337366] A review of the system's log file was conducted, which confirmed that there were no malfunctions observed during the procedure. No system malfunctions were observed. A review of the device history record (dhr) for the aquabeam robotic system, lot number 19c00267, was conducted, which confirmed that there were no nonconformances generated during the manufacturing process of this system, which could relate to the reported event. The review indicated that the system met all required specifications when released for distribution. A review for similar events was conducted on the aquabeam robotic system, lot number 19c00267, which confirmed that there were no other similar events reported on this system. The aquabeam system's instructions for use (ifu), ifu0101, rev. C, was reviewed and bleeding is listed as a potential perioperative risk of the aquablation procedure. The system was not returned for investigation of this complaint. Bleeding is a potential risk of the aquablation procedure. Based on the review of the log file, dhr, and ifu, the event is considered not to be device related.
Patient Sequence No: 1, Text Type: N, H10


[188337367] A male patient underwent an aquablation procedure. Per standard post-aquablation procedure, focal cauterization was performed, and the surgeon decided to administer tranexamic acid (txa) to address bleeding and the patient was transferred to the post-anesthesia care unit (pacu). The patient was later taken back to the or for persistent bleeding (per manufacturer's instructions for use, bleeding is a potential perioperative risk of the aquablation procedure). The patient presented with an arterial bleeder at the bladder neck as well as an untreated median lobe. Cauterization of the area was performed to stop the bleeding. The patient received two (2) units of blood transfusion and discharged home two (2) days post-aquablation procedure without any clinical sequela. No malfunction of the aquabeam robotic system was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012977056-2020-00010
MDR Report Key9803073
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-06
Date of Report2020-03-06
Date of Event2020-02-05
Date Mfgr Received2020-02-07
Device Manufacturer Date2019-03-08
Date Added to Maude2020-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DORIA ESQUIVEL
Manufacturer Street900 ISLAND DRIVE SUITE 170
Manufacturer CityREDWOOD CITY, CA
Manufacturer CountryUS
Manufacturer Phone2327291
Manufacturer G1PROCEPT BIOROBOTICS CORPORATION
Manufacturer Street900 ISLAND DRIVE SUITE 101
Manufacturer CityREDWOOD CITY, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAQUABEAM ROBOTIC SYSTEM
Generic NameFLUID JET REMOVAL SYSTEM
Product CodePZP
Date Received2020-03-06
Model NumberAB2000
Catalog NumberAB2000
Lot Number19C00267
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPROCEPT BIOROBOTICS CORPORATION
Manufacturer Address900 ISLAND DRIVE SUITE 101 REDWOOD CITY, CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.