ICEROD CX 90? NEEDLE/VL FPRPR3533

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-08 for ICEROD CX 90? NEEDLE/VL FPRPR3533 manufactured by Galil Medical Ltd.

Event Text Entries

[182560233] Two icerod cx 90 needles were used in a renal cryoablation procedure. The cryoablation system and needles both passed testing with no issues. The patient experienced muscle twitching during the first freeze cycle. The user moved the needles to different channels and identified the needle a7116-025 as the needle causing the twitching. A new needle was used to continue with the procedure, and performed as expected with no issues. The case was completed with no injury reported for the patient. The defective needle was discarded and will not be returned to the manufacturer for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616793-2020-00056
MDR Report Key9803844
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-08
Date of Report2020-03-08
Date of Event2020-02-07
Date Mfgr Received2020-02-07
Date Added to Maude2020-03-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SANDRA BAUSBACK-ABALLO
Manufacturer Street300 CONSHOHOCKEN STATE RD 300 FOUR FALLS CORPORATE CTR
Manufacturer CityWEST CONSHOHOCKEN, PA
Manufacturer CountryUS
Manufacturer Phone3311537
Manufacturer G1GALIL MEDICAL LTD.
Manufacturer Street1 TAVOR BUILDING PO BOX 224, INDUSTRIAL PARK
Manufacturer CityYOKNEAM, 2069203,
Manufacturer CountryIS
Manufacturer Postal Code2069203,
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICEROD CX 90? NEEDLE/VL
Generic NameCRYOABLATION NEEDLE
Product CodeGEH
Date Received2020-03-08
Model NumberFPRPR3533
Catalog NumberFPRPR3533
Lot NumberA7116
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGALIL MEDICAL LTD
Manufacturer Address1 TAVOR BUILDING P.O. BOX 224, INDUSTRIAL PARK YOKNEAM, 2069203, IS 2069203,


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-08