TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM PN-004 075 PN-004075

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-09 for TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM PN-004 075 PN-004075 manufactured by Venusa De Mexico S.a. De C.v..

Event Text Entries

[183749319] One tacticath quartz contact force ablation catheter was received for evaluation. The device was returned due to noise. Multiple short circuits were noted, consistent with the reported noise. Fluid was noted outside of the irrigation tubing but within the tygon tubing proximal to the catheter handle. In addition, dried saline was noted within the connector plug, consistent with the short circuits detected. Further testing revealed a leak at the luer/irrigation tubing junction due to a gap in the adhesive between the irrigation tube and the proximal tube inside the luer lock, allowing fluid to flow outside of the irrigation tubing and into the proximal tubing towards the optical and electrical connectors, consistent with the dried saline noted within connector plug.
Patient Sequence No: 1, Text Type: N, H10


[183749320] This report is to advise of an event observed during analysis confirming a tygon tubing leak and short circuit of the catheter.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680001-2020-00012
MDR Report Key9804953
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-09
Date of Report2020-03-09
Date of Event2020-01-09
Date Mfgr Received2020-02-29
Device Manufacturer Date2019-07-29
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1VENUSA DE MEXICO S.A. DE C.V.
Manufacturer StreetCALLE HERTZ 1525-6 PARQUE INDUSTRIAL J. BERMUDEZ
Manufacturer CityCIUDAD JUAREZ, CHIHUAHUA 32470
Manufacturer CountryMX
Manufacturer Postal Code32470
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Generic NameCATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Product CodeOAE
Date Received2020-03-09
Returned To Mfg2020-02-06
Model NumberPN-004 075
Catalog NumberPN-004075
Lot Number6884772
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerVENUSA DE MEXICO S.A. DE C.V.
Manufacturer AddressCALLE HERTZ 1525-6 PARQUE INDUSTRIAL J. BERMUDEZ CIUDAD JUAREZ, CHIHUAHUA 32470 MX 32470


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.