HF SENSOR DELIVERY SYSTEM CM2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-09 for HF SENSOR DELIVERY SYSTEM CM2000 manufactured by St. Jude Medical, Inc..

Event Text Entries

[182523748] The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
Patient Sequence No: 1, Text Type: N, H10


[182523749] On (b)(6) 2020 a right heart catheterization was performed to confirm the validity of the pressure sensor readings. The sensor was recalibrated, and the mean was increased by 9. 8 mmhg. Accurate readings were observed under the manufacturer's patient care network database.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004936110-2020-00115
MDR Report Key9804963
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-09
Date of Report2020-03-09
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, INC.
Manufacturer Street387 TECHNOLOGY CIRCLE NW SUITE 500
Manufacturer CityATLANTA GA 30313
Manufacturer CountryUS
Manufacturer Postal Code30313
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHF SENSOR DELIVERY SYSTEM
Generic NameSYSTEM, HEMODYNAMIC, IMPLANTABLE
Product CodeMOM
Date Received2020-03-09
Model NumberCM2000
Catalog NumberCM2000
Lot Number6552436
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.
Manufacturer Address387 TECHNOLOGY CIRCLE NW SUITE 500 ATLANTA GA 30313 US 30313


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.