SMITHS MEDICAL LEVEL 1 HOTLINE TRAUMA FAST FLOW SYSTEMS 8002950

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-09 for SMITHS MEDICAL LEVEL 1 HOTLINE TRAUMA FAST FLOW SYSTEMS 8002950 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[182562317] One fluid warmer was returned for evaluation. Visual inspection of the device found it to be in fair condition, with air detector door removed by customer. Device underwent functional testing by attaching a f-30 gas/vent filter to air detector and powered it on. The air detector clamp test was completed and the plunger retainer was noted to be stuck halfway and then would clamp all the way down. This confirms the reported customer allegation, as a result of faulty solenoid. The problem source has been determined to be unknown. While no definitive problem source to the reported issue could be determined, this investigation revealed no intrinsic evidence to suggest a cause of issue related to manufacturing.
Patient Sequence No: 1, Text Type: N, H10


[182562318] Information was received that the clamp on a smiths medical level 1 trauma fast flow system will not work its way down when the cassette is removed. No adverse effects reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-01900
MDR Report Key9805888
Report SourceUSER FACILITY
Date Received2020-03-09
Date of Report2020-03-09
Date Mfgr Received2020-02-11
Device Manufacturer Date2007-04-09
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMITHS MEDICAL LEVEL 1 HOTLINE TRAUMA FAST FLOW SYSTEMS
Generic NameWARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Product CodeBSB
Date Received2020-03-09
Returned To Mfg2019-10-02
Model Number8002950
Catalog Number8002950
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE N MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.