O-ARM 1000 IMAGING SYSTEM 9732719

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-09 for O-ARM 1000 IMAGING SYSTEM 9732719 manufactured by Medtronic Navigation, Inc (littleton).

Event Text Entries

[183280834] Device evaluation has not been done at the time of submitting this report. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[183280835] Medtronic received information regarding an imaging system used during a sacroiliac and thoracolumbar procedure. It was reported that they felt inaccurate. When navigating on the navigation system, they felt they were "too deep" when they should have been more superficial (length was unknown). They took another spin and felt they were still inaccurate and confirmed this by taking an ap and lateral shot using a different system, they proceeded with the case by switching out the reference frame and the system. There was a procedure delay of one hour or longer and no impact to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004785967-2020-00339
MDR Report Key9806156
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-09
Date of Report2020-03-25
Date of Event2020-03-05
Date Mfgr Received2020-03-09
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Street300 FOSTER ST
Manufacturer CityLITTLETON MA 01460
Manufacturer CountryUS
Manufacturer Postal Code01460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO-ARM 1000 IMAGING SYSTEM
Generic NameIMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Product CodeOXO
Date Received2020-03-09
Model Number9732719
Catalog Number9732719
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Address300 FOSTER ST LITTLETON MA 01460 US 01460


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-09