OPTICROSS 8655

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-09 for OPTICROSS 8655 manufactured by Boston Scientific Corporation.

Event Text Entries

[182727846] The device was returned for analysis. Visual inspection revealed a detachment was observed of the imaging window from the lapjoint section. Imaging core windup began 23. 60cm from the distal end of the connector shaft. A kink was observed in the sheath assembly from the femoral marker to the proximal end. No other visual damages were encountered upon visual inspection. Impedance testing showed an electrical open at the proximal wave form. The distal housing and the transducer assembly appeared to be in good conditions. The lapjoint measure was within specification.
Patient Sequence No: 1, Text Type: N, H10


[182727850] Reportable based on device analysis completed on 18feb2020. It was reported that lost image occurred. An opticross imaging catheter was prepped for use without issue. During the procedure, however, the screen became black, and nothing was displayed. The procedure was completed with another of same device. No patient complications were reported. However, returned device analysis revealed a detachment of the imaging window from the the lapjoint section.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02817
MDR Report Key9808532
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-09
Date of Report2020-03-09
Date of Event2020-01-15
Date Mfgr Received2020-02-18
Device Manufacturer Date2019-09-06
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street2546 CALLE PRIMERA
Manufacturer CityALAJUELA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTICROSS
Generic NameCATHETER, ULTRASOUND, INTRAVASCULAR
Product CodeOBJ
Date Received2020-03-09
Returned To Mfg2020-01-22
Model Number8655
Catalog Number8655
Lot Number0024388312
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.