ZOLL IVTM THERMOGARD XP 8700-0650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-09 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.

Event Text Entries

[188220751] An onsite evaluation of the thermogard xp ivtm console (sn (b)(4)) was performed by a zoll service technician. The reported complaint of a tcmid: 05 (coolant diff failure) was not observed during functional testing; however was confirmed during review of the event log. The probable root cause was due to the damaged lm 34 temperature sensor and damaged pic 16 board likely due to normal wear and tear of the console. The thermogard xp ivtm console is a reusable device and was manufactured on 30 october 2011. It has exceeded its expected serviceable life of five years and is over 8 years old. Upon visual inspection no physical damage was observed. A review of the event log was performed and several tcmid: 05 error were observed. After replacing the lm 34 temperature sensor and pic 16 board, the console passed all testing criteria and meets performance specifications. Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for thermogard xp ivtm console with serial number (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188220752] During device check, the thermogard console (sn (b)(4)) displayed tcmid: 05 (coolant diff failure). No patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00232
MDR Report Key9808770
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-09
Date of Report2020-03-09
Date Mfgr Received2020-02-14
Device Manufacturer Date2011-11-30
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM THERMOGARD XP
Generic NameTHERMAL REGULATING SYSTEM
Product CodeNCX
Date Received2020-03-09
Model Number8700-0650
Catalog Number8700-0650
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-09

© 2024 FDA.report