EXTENSION, 30CM 3383

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-09 for EXTENSION, 30CM 3383 manufactured by St. Jude Medical - Neuromodulation.

Event Text Entries

[182699555] The results/method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10


[182699556] Related manufacturer reference number: 1627487-2020-02552; related manufacturer reference number: 1627487-2020-02554; related manufacturer reference number: 3006705815-2020-01089. It was reported that the patient? S stimulation kept stopping on its own. Additionally, patient reported sometimes experiencing electrical sensations at the lead/extension connection site. Diagnostics revealed high impedances. X-ray revealed that one of the lead tails was not completely inserted into the extension. Reprogramming was unable to resolve the issue. As such, surgical intervention took place on (b)(6) 2020. During intra-op, impedance reading was normal on the lead and extension. However, the impedance reading was high when measured on the entire system. The extensions were replaced with new extensions, but the issue persisted. In turn, the ipg was also explanted and replaced with a ipg to address the issue. No intervention took place on the lead.
Patient Sequence No: 1, Text Type: D, B5


[182809258] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[182809259] Additional information received indicates that the physician experienced difficulty inserting the new extensions into the ipg and the high impedance persisted. In turn, the ipg was also explanted and replaced with a ipg resolving the issue. Reportedly therapy was resumed.
Patient Sequence No: 1, Text Type: D, B5


[187480209] The reported event for intermittent stimulation was not confirmed. At receipt, both extension lead have no anomaly and passed functional testing. No root cause was identified for the reported issue
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1627487-2020-02553
MDR Report Key9809127
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-09
Date of Report2020-03-10
Date of Event2019-12-16
Date Mfgr Received2020-03-04
Device Manufacturer Date2016-09-13
Date Added to Maude2020-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANDREA DEITZ
Manufacturer Street6901 PRESTON ROAD
Manufacturer CityPLANO TX 75024
Manufacturer CountryUS
Manufacturer Postal75024
Manufacturer Phone9723098000
Manufacturer G1ST. JUDE MEDICAL - NEUROMODULATION
Manufacturer Street6901 PRESTON RD
Manufacturer CityPLANO TX 75024
Manufacturer CountryUS
Manufacturer Postal Code75024
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXTENSION, 30CM
Generic NameSCS EXTENSION
Product CodeGZB
Date Received2020-03-09
Model Number3383
Catalog Number3383
Lot Number5637822
Device Expiration Date2018-09-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL - NEUROMODULATION
Manufacturer Address6901 PRESTON RD PLANO TX 75024 US 75024


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.