PRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT G1460J

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-01-17 for PRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT G1460J manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[17371955] User alleges one event of when performed a blood draw and, when went to engage the needle, the needle could not click into the needle protection device. Needle stick reported. No treatment.
Patient Sequence No: 1, Text Type: D, B5


[17543577] Results evaluation (other): our investigation into this event is very limited as the user facility did not return the device used in this event. A review of our complaints database show no other reports on this device lot number of the finished good or the needle lot number. A further review shows no similar reports received on this catalog number. A review of the mfg and incoming records show no problems noted that would explain this event. The mfg lot size, for the reported lot number, was 53,300 units. Without the device sample we are unable to determine the exact cause of this event. Due to no similar reports received, we feel this is an isolated incident and it has been logged for trending.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217052-2008-00015
MDR Report Key981141
Report Source07
Date Received2008-01-17
Date of Report2007-12-26
Date of Event2007-12-25
Date Mfgr Received2007-12-26
Device Manufacturer Date2007-05-01
Date Added to Maude2008-07-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactTIMOTHY MATSON
Manufacturer Street10 BOWMAN DR.
Manufacturer CityKEENE NH 03431
Manufacturer CountryUS
Manufacturer Postal03431
Manufacturer Phone6033523812
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer CityKEENE NH 03431
Manufacturer CountryUS
Manufacturer Postal Code03431
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT
Generic Name75 GJE- TRAY, BLOOD COLLECTION
Product CodeGJE
Date Received2008-01-17
Catalog NumberG1460J
Lot Number1157891
Device Expiration Date2010-04-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key985109
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address10 BOWMAN DR. KEENE NH 03431 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-17

© 2024 FDA.report