CIDEX? OPA SOLUTION 20390

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-10 for CIDEX? OPA SOLUTION 20390 manufactured by Advanced Sterilization Products.

Event Text Entries

[186101309] Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186101310] A customer reported using cidex? Opa solution past the 14-day reuse date and the instruments were released and used on patients. The customer stated the solution was tested prior to use and met the minimum effective concentration (mec). There were no reports of any injury, harm or infections reported. Per the instructions for use (ifu), cidex? Opa solution may be reused for up to a maximum of 14 days provided the required conditions of ortho-phthalaldehyde concentration and temperature exist based upon monitoring described in the directions for use. Do not rely solely on days in use. Concentration of this product during its reuse life must be verified by the cidex? Opa solution test strip prior to each use to determine that the concentration of ortho-phthalaldehyde is above the mec of 0. 3%. The product must be discarded after 14 days, even if the cidex? Opa solution test strip indicates a concentration above the mec. Although there is no patient harm or injury reported, advanced sterilization products has decided to report cases when the customer uses expired product and releases the instruments for use since high-level disinfection cannot be assured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2020-50028
MDR Report Key9812390
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-10
Date of Report2020-03-10
Date of Event2020-02-12
Date Mfgr Received2020-02-12
Device Manufacturer Date2000-01-01
Date Added to Maude2020-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGABRIELA MCLELLAN
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9495030264
Manufacturer G1ASP IRVINE MFG
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCIDEX? OPA SOLUTION
Generic NameBIOCIDES SOLUTIONS (MED)
Product CodeMED
Date Received2020-03-10
Catalog Number20390
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-10