STERRAD? 100NX STERILIZER 1-DR 10104

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-10 for STERRAD? 100NX STERILIZER 1-DR 10104 manufactured by Advanced Sterilization Products.

Event Text Entries

[186100992] A field service engineer was dispatched to the customer site. The catalytic converter, oil mist filter, and vacuum pump oil were replaced to resolve the fog/haze issue. Unit meets specifications and was returned to service. Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186100993] A customer reported a fog or haze emitting from the sterrad? 100nx sterilizer. There was no report of any injuries or human reactions. The customer powered down the unit and discontinued its use. An asp field service engineer was dispatched to assess the unit onsite. This event is being reported as a malfunction report subsequent to a serious injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2020-50020
MDR Report Key9812437
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-10
Date of Report2020-03-10
Date of Event2020-02-12
Date Mfgr Received2020-02-12
Device Manufacturer Date2000-01-01
Date Added to Maude2020-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. GABRIELA MCLELLAN
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9495030264
Manufacturer G1ASP IRVINE MFG
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERRAD? 100NX STERILIZER 1-DR
Generic NameSTERRAD EQUIPMENT
Product CodeMLR
Date Received2020-03-10
Model Number10104
Catalog Number10104
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.