PCS?2 PLASMA COLLECTION SYSTEM 06002-110-NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-10 for PCS?2 PLASMA COLLECTION SYSTEM 06002-110-NA manufactured by Haemonetics Corporation.

Event Text Entries

[182768076] There was no adverse reaction or injury experienced during the donation procedure, the donor was reported to have left the center in good health. A haemonetics field service engineer evaluated the suspected device and found there were no faults observed which could negatively impact the donor.
Patient Sequence No: 1, Text Type: N, H10


[182768077] On october 25 2018 haemonetics was notified of a donor fatality which had occurred within 24 hours of the donor's last plasma donation procedure, utilizing the pcs? 2 plasma collection system. The plasma collection procedure was completed successfully, with no error messages or alarms encountered. The pcs? 2 plasma collection system collected the target volume of 876ml of plasma and 500ml of saline administered per the routine procedure. The cause of death was reported to be not known, a copy of the coroner's report was not available at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219343-2018-00052
MDR Report Key9812720
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-10
Date of Report2018-10-25
Date of Event2018-10-24
Date Mfgr Received2018-10-25
Device Manufacturer Date2007-09-28
Date Added to Maude2020-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JANICE KISER
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Manufacturer G1HAEMONETICS CORPORATION
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePCS?2 PLASMA COLLECTION SYSTEM
Generic NamePCS2,LIST NO. 06002-110-NA
Product CodeGKT
Date Received2020-03-10
Model Number06002-110-NA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION
Manufacturer Address125 SUMMER STREET BOSTON, MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.