ZIMMER SKIN GRAFT MESHER 00770100000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-10 for ZIMMER SKIN GRAFT MESHER 00770100000 manufactured by Zimmer Surgical, Inc..

Event Text Entries

[183916899] This event has been recorded by zimmer biomet under (b)(4). Udi #: (b)(4). The device history record and previous repair record for zimmer skin graft mesher serial number (b)(4) were reviewed and noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with the device. The record reviews found no issues with the device and all verifications, inspections, and tests were successfully completed. The reported event was confirmed by the service technician who performed the evaluation and repair. On 10 december 2019, it was reported that a mesher needed to be repaired. A zimmer skin graft mesher serial number (b)(4) was already at a zimmer facility and was available for evaluation. Evaluation of the device on (b)(6) 2019 found that the comb was bent and that none of the pretests were able to be performed. Repair of the mesher occurred the same day and involved replacing the comb. The technician then recalibrated the device and verified that it was functioning as intended. The mesher was then returned to the customer without further incident. The device was tested, inspected, and repaired. While the service technician confirmed that the comb was bent, it cannot be determined from the information provided as to what caused the comb to become bent. Therefore, the root cause of the reported event cannot be determined. The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process. Review of the information provided during the investigation determined that there are no further actions needed at this time. This complaint will be tracked and trended for any adverse trends that may warrant further action.
Patient Sequence No: 1, Text Type: N, H10


[183916900] It was reported that a returning loaner device required repair. During an evaluation of the device, it was discovered that there was a bent comb. No adverse events were reported as a result of this malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001526350-2020-00272
MDR Report Key9813001
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-10
Date of Report2020-03-10
Date of Event2019-12-10
Date Mfgr Received2020-03-10
Device Manufacturer Date2015-11-09
Date Added to Maude2020-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER SURGICAL, INC.
Manufacturer Street200 WEST OHIO AVENUE
Manufacturer CityDOVER OH 44622
Manufacturer CountryUS
Manufacturer Postal Code44622
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameZIMMER SKIN GRAFT MESHER
Generic NameEXPANDER, SURGICAL, SKIN GRAFT
Product CodeFZW
Date Received2020-03-10
Returned To Mfg2019-12-10
Catalog Number00770100000
Lot Number63196962
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER SURGICAL, INC.
Manufacturer Address200 WEST OHIO AVENUE DOVER OH 44622 US 44622


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-10

© 2026 FDA.report