XPS? SYSTEM NAVIGATED BLADE 1884012EM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-10 for XPS? SYSTEM NAVIGATED BLADE 1884012EM manufactured by Medtronic Xomed Inc..

Event Text Entries

[182996233] Product analysis found out that the tracker hub became detached from the outer hub which would have resulted in the reported event. The component that became detached is not likely to become un-retrieved device fragment. There were no signs of misuse or mishandling. The tracker hub shall be bonded to the outer hub using adhesive per the manufacturing instructions. When viewed under magnification, there was a residue consistent with adhesive on the mounting surfaces. Although it appears that adhesive was present, the information indicates there was insufficient adhesion which resulted in the component(s) coming loose. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[182996314] A health care provider (hcp) reported that when opening the product, the em part came off. The inner cylinder part protruded during a ess (endoscopic sinus surgery). There was no patient impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00148
MDR Report Key9813748
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-10
Date of Report2020-03-10
Date of Event2020-02-13
Date Mfgr Received2020-02-13
Date Added to Maude2020-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXPS? SYSTEM NAVIGATED BLADE
Generic NameBUR, EAR, NOSE AND THROAT
Product CodeEQJ
Date Received2020-03-10
Returned To Mfg2020-02-18
Model Number1884012EM
Catalog Number1884012EM
Lot Number0218440969
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.