ICEROD CX 90? NEEDLE/VL FPRPR3533

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-11 for ICEROD CX 90? NEEDLE/VL FPRPR3533 manufactured by Galil Medical Ltd.

Event Text Entries

[183175399] The needle was returned to the manufacturer for investigation. The needle was found to be bent near the handle. The shaft temperature was found to be below specifications. Ice formed along the entire sleeve, confirming the reported complaint. Disassembly of the needle revealed that the vacuum sleeve had collapsed along the entire needle shaft. No soldering gaps, soldering flux or corrosive signs were visible on the sleeve's external surfaces. Multiple possible root causes were identified that could have caused the malfunction: vacuum sleeve thermal insulation loss, bending of the shaft or vacuum sleeve collapse onto the needle. The needle lot manufacturing records were reviewed, and no related deviations were found within the needle batch.
Patient Sequence No: 1, Text Type: N, H10


[183175400] Two needles were used in a renal cryoablation procedure. During the freeze cycle, one of the needle's collected frost on the needle shaft. The patient's skin was protected by the user to continue the case and prevent injury. The iceball size was satisfactory for the user. The case was completed with no reported injury to the patient. The reported defective needle was returned to the manufacturer for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616793-2020-00059
MDR Report Key9817123
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2020-02-11
Date Mfgr Received2020-02-11
Device Manufacturer Date2019-07-10
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SANDRA BAUSBACK-ABALLO
Manufacturer Street300 CONSHOHOCKEN STATE RD 300 FOUR FALLS CORPORATE CTR
Manufacturer CityWEST CONSHOHOCKEN, PA
Manufacturer CountryUS
Manufacturer Phone3311537
Manufacturer G1GALIL MEDICAL LTD.
Manufacturer Street1 TAVOR BUILDING PO BOX 224, INDUSTRIAL PARK
Manufacturer CityYOKNEAM, 2069203
Manufacturer CountryIS
Manufacturer Postal Code2069203
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICEROD CX 90? NEEDLE/VL
Generic NameCRYOABLATION NEEDLE
Product CodeGEH
Date Received2020-03-11
Returned To Mfg2020-02-23
Model NumberFPRPR3533
Catalog NumberFPRPR3533
Lot NumberA6961
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGALIL MEDICAL LTD
Manufacturer Address1 TAVOR BUILDING P.O. BOX 224 YOKNEAM, 2069203 IS 2069203


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2026 FDA.report