[188112851] The complaint notification associated with this device described that one screw in the knee joint is loose. The user did not fall, no serious injuries were sustained as a result of the failure and no medical treatment was required. The evaluation of this specific device was conducted at the manufacturer's after sales service center on january 10th, 2018. After evaluation we found that all bolts including the upper posterior section of the knee joint where loose. An exact reason for the loosening of the bolt could not be determined. In this specific case the failure did not lead to a fall and/or serious injury, but it could cause or contribute to a serious injury if it were to recur. Therefore, we report this failure according to guidance for industry and fda staff on "mdr for manufacturers" chapter 2. 15.
Patient Sequence No: 1, Text Type: N, H10


[188112852] Knee joint was sent in for repair. According to information provided by the customer the bolts were loose. Patient fell, but there were no serious injuries sustained as a result of the fall.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616494-2018-00053
MDR Report Key9817124
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-11
Date of Report2018-01-25
Date Mfgr Received2017-11-17
Device Manufacturer Date2015-10-07
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DOMINIK SCHMIDT
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, 37115
Manufacturer CountryGM
Manufacturer Postal37115
Manufacturer G1OTTO BOCK HEALTHCARE GMBH
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, 37115
Manufacturer CountryGM
Manufacturer Postal Code37115
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMODULAR POLYCENTRIC EBS KNEE JOINT K/D
Generic NameMODULAR POLYCENTRIC EBS KNEE JOINT K/D
Product CodeISY
Date Received2020-03-11
Returned To Mfg2018-01-02
Model Number3R60=KD
Catalog Number3R60=KD
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE GMBH
Manufacturer AddressMAX-NAEDER-STR. 15 DUDERSTADT, 37115 GM 37115


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11