MODULAR EBS KNEE JOINT 3R60

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-11 for MODULAR EBS KNEE JOINT 3R60 manufactured by Otto Bock Healthcare Gmbh.

Event Text Entries

[188111933] Customer stated that 4 bolts in the knee came loose, 2 bolts fell out completely. Client was involved in a motorcycle accident (b)(6) 2017. The customer indicated the device (3r60) was the cause of the fall. The client did seek medical attention and was hospitalized with multiple fractures. The evaluation of this specific device was conducted at the manufacturer's after sales service center on january 29th, 2018. Due to the time difference between the accident and the technical evaluation we can not identify the original failure which might have led to the fall. An exact reason for the loosening of the bolt could not be determined. We are also not able to assess the relationship between the event and the product based on the available information. Therefore, we report this failure according to guidance for industry and fda staff on "mdr for manufacturers" chapter 2. 15.
Patient Sequence No: 1, Text Type: N, H10


[188111934] Knee joint was sent in for repair. Customer stated that 4 bolts in the knee came loose, 2 bolts fell out completely. Client was involved in a motorcycle accident (b)(6) 2017. The customer indicated the device (3r60) was the cause of the fall. The client did seek medical attention and was hospitalized with multiple fractures. The accident happened in (b)(6) 2017. The joint was not sent in for evaluation until (b)(6) 2018. We report out of an abundance of caution.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616494-2018-00054
MDR Report Key9817132
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-11
Date of Report2018-01-31
Date of Event2017-07-15
Date Mfgr Received2018-01-16
Device Manufacturer Date2015-10-07
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DOMINIK SCHMIDT
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, 37115
Manufacturer CountryGM
Manufacturer Postal37115
Manufacturer G1OTTO BOCK HEALTHCARE GMBH
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, 37115
Manufacturer CountryGM
Manufacturer Postal Code37115
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMODULAR EBS KNEE JOINT
Generic NameMODULAR EBS KNEE JOINT
Product CodeISY
Date Received2020-03-11
Returned To Mfg2018-01-29
Model Number3R60
Catalog Number3R60
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE GMBH
Manufacturer AddressMAX-NAEDER-STR. 15 DUDERSTADT, 37115 GM 37115


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2020-03-11

© 2026 FDA.report