[188112156] Product and/or additional information not arrived until now although requested several times. No information provided by the customer if there were serious injuries sustained as a result of the failure. A specific investigation could not be conducted until now. The complaint cannot be reproduced at this time.
Patient Sequence No: 1, Text Type: N, H10


[188112157] The lock lever fell into place while the patient (who is also the practitioner) was trying to clear a step. The locked knee did not allow him get this foot all the way over the step, he hit the step with his toes and fell. This did not completely solve the problem, the patient is now in the hospital due to a fall.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616494-2018-00064
MDR Report Key9817248
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-11
Date of Report2018-05-04
Date of Event2018-04-03
Date Mfgr Received2018-04-03
Device Manufacturer Date2017-02-09
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DOMINIK SCHMIDT
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, NIEDERSACHSEN 37115
Manufacturer CountryGM
Manufacturer Postal37115
Manufacturer G1OTTO BOCK HEALTHCARE GMBH
Manufacturer StreetMAX-NAEDER-STR. 15
Manufacturer CityDUDERSTADT, 37115
Manufacturer CountryGM
Manufacturer Postal Code37115
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMODULAR KNEE JOINT WITH ROTARY HYDRAULIC
Generic NameMODULAR KNEE JOINT WITH ROTARY HYDRAULIC
Product CodeISY
Date Received2020-03-11
Model Number3R80
Catalog Number3R80
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE GMBH
Manufacturer AddressMAX-NAEDER-STR. 15 DUDERSTADT, 37115 GM 37115


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11