INTRALASE FS2 20003D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-11 for INTRALASE FS2 20003D manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[182984143] (b)(4). A review of records related to the device including labeling, complaint trending, and risk documentation will be performed. Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[182984144] It was reported that a laser vision correction patient had surgery on (b)(6) 2020 and presented on (b)(6) 2020 with ingrowth in the left eye (os) post treatment. A flap lift and rinse was performed on (b)(6) 2020. It was stated that the patient had no loss of best corrected visual acuity (bcva). The patient complained of blurry vision. Patient reported symptoms are not interfering with daily activities. It was reported that symptoms resolved on (b)(6) 2020 and patient is doing better. Bcva from (b)(6) 2020: right eye pre-op 20/25 1. 75 x -1. 00 x 60, left eye pre-op 20/20 2. 00 x -. 75 x 155. Bcva from (b)(6) 2020: right eye post-op 20/20. 50 x -. 75 x 170, left eye post-op 20/20 -1. 50 x -2. 50 x 135.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006695864-2020-00168
MDR Report Key9817553
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2020-02-10
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-17
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Street510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameINTRALASE FS2
Generic NameFEMTOSECOND LASER
Product CodeHNO
Date Received2020-03-11
Model Number20003D
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E ST ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-11
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-11