30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT 80395

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-11 for 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT 80395 manufactured by Dentsply Llc.

Event Text Entries

[186735105] There has been a previous report received where lack of water flow has caused an overheating insert. Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur. As such, this event meets the criteria for reportability per 21 cfr part 803. The device is available for evaluation, though has not been returned as of this report. Evaluation results will be submitted as they become available.
Patient Sequence No: 1, Text Type: N, H10


[186735106] While using a 30k fsi-sli-10s insert, it did not dispense water and the insert was getting warm; no injury resulted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2424472-2020-00014
MDR Report Key9817717
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-11
Date of Report2020-03-05
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK, PA
Manufacturer CountryUS
Manufacturer Phone8494229
Manufacturer G1DENTSPLY LLC
Manufacturer Street1301 SMILE WAY
Manufacturer CityYORK, PA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT
Generic NameSCALER, ULTRASONIC
Product CodeELC
Date Received2020-03-11
Model NumberNA
Catalog Number80395
Lot Number18017/HW
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY LLC
Manufacturer Address1301 SMILE WAY YORK, PA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.