RAPIDLAB 1265 BLOOD GAS ANALYZER 10321852

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-11 for RAPIDLAB 1265 BLOOD GAS ANALYZER 10321852 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[182930880] The customer stated that repeat testing was performed to confirm correct results which met the clinical condition of the patient. Siemens service communicated with the customer the following morning after the event to assess instrument performance. The customer had replaced the lactate sensor. A field service engineer (fse) went onsite to export log files and gather additional information. According to the fse, the high gl and qc results were in the expected ranges after the lactate sensor replacement. Siemens has requested the original lactate sensor to be returned but it is not available, and the log files for investigation.
Patient Sequence No: 1, Text Type: N, H10


[182930881] The customer reported discrepant low lactate results from one patient on the rl 1265 when compared to the lactate results on the rp 500. The customer stated the patient was in critical condition and admitted to the icu for several days prior to the event. The patient had sepsis which rapidly progressed. The customer indicated the patient is deceased. Safety data information from the complaint states that the customer in not alleging harm to the patient due to the use of the device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002637618-2020-00009
MDR Report Key9817843
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2020-03-02
Date Mfgr Received2020-03-06
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFELIX AKINRINOLA
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD, MA
Manufacturer CountryUS
Manufacturer Phone7052212
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY, CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRAPIDLAB 1265 BLOOD GAS ANALYZER
Generic NameRL 1265
Product CodeCHL
Date Received2020-03-11
Model Number10321852
Catalog Number10321852
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN, NY US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11