JETSTREAM XC 45007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-11 for JETSTREAM XC 45007 manufactured by Boston Scientific Corporation.

Event Text Entries

[183100509] Age at time of event: patient is over 18 years old.
Patient Sequence No: 1, Text Type: N, H10


[183100510] It was reported that the device stopped spinning. A jetstream xc catheter, 2. 4mm was selected for a lower extremity atherectomy procedure. During the procedure, the device stopped spinning while engaged in the plaque. The sound abruptly changed when the issue occurred. The device was safely removed and the procedure was completed successfully by a balloon angioplasty and stent placement. No patient complications were reported and there were no adverse effects regarding the patient. The patient was fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02859
MDR Report Key9818392
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2020-02-19
Date Mfgr Received2020-02-19
Device Manufacturer Date2019-10-09
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetMODEL FARM ROAD
Manufacturer CityCORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJETSTREAM XC
Generic NameCATHETER, PERIPHERAL, ATHERECTOMY
Product CodeMCW
Date Received2020-03-11
Model Number45007
Catalog Number45007
Lot Number0024570151
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.