CYCLESURE? 24 BIO INDICATOR 14324

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-11 for CYCLESURE? 24 BIO INDICATOR 14324 manufactured by Advanced Sterilization Products.

Event Text Entries

[186101694] Concomitant medical product: sterrad 100s sterilizer, serial #: unknown. Initial reporter phone #: (b)(6). (b)(4). Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186101695] A customer reported a positive result with a cyclesure? 24 biological indicator (bi) after a completed sterrad? 100s cycle. The bi was incubated for 24 hours. The chemical indicator (ci) changed color correctly. The previous and subsequent bi results were both negative. The affected load was partially released and used on a patient. There was no report of infection, injury or harm to patient(s) associated with this issue. Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured. Therefore, as a matter of policy asp had decided to report all incidents of positive cyclesure? 24 biological indicators when the load has been released and used on patient(s) prior to reprocessing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2020-00021
MDR Report Key9818853
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-11
Date of Report2020-02-21
Date of Event2020-01-02
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. GABRIELA MCLELLAN
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone949453-639
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCYCLESURE? 24 BIO INDICATOR
Generic NameINDICATOR, BIOLOGICAL
Product CodeFRC
Date Received2020-03-11
Model Number14324
Catalog Number14324
Lot Number16819029
Device Expiration Date2020-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.