MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-11 for INDIGO SYSTEM ASPIRATION CATHETER 6 CAT6KIT-A CAT6KIT manufactured by Penumbra, Inc..
[182964023]
Please note that this complaint was submitted to the fda by the user facility with the following reference number: mw5093209. The product was not returned for evaluation. Without the return of the device, the root cause of the problem cannot be determined. The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
Patient Sequence No: 1, Text Type: N, H10
[182964024]
The patient was undergoing a thrombectomy procedure in the left superficial femoral artery (sfa) using an indigo system aspiration catheter 6 (cat6). During the procedure, the cat6 became stuck in the vessel between a heavily calcified area and the subintimal space. When the physician attempted to pull the cat6 out of the patient, a small piece of the shaft broke off inside the vessel. Subsequently, a goose neck snare was used to successfully remove the piece of cat6. The procedure was completed using another cat6. There was no report of an adverse effect to the patient.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005168196-2020-00327 |
MDR Report Key | 9818916 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-11 |
Date of Report | 2020-02-10 |
Date of Event | 2020-02-10 |
Report Date | 2005-01-01 |
Date Reported to FDA | 2005-01-01 |
Date Reported to Mfgr | 2005-01-10 |
Date Mfgr Received | 2020-02-10 |
Date Added to Maude | 2020-03-11 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. VERONICA FARRIS |
Manufacturer Street | ONE PENUMBRA PLACE |
Manufacturer City | ALAMEDA CA 94502 |
Manufacturer Country | US |
Manufacturer Postal | 94502 |
Manufacturer Phone | 5107483200 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | INDIGO SYSTEM ASPIRATION CATHETER 6 |
Generic Name | QEW |
Product Code | QEW |
Date Received | 2020-03-11 |
Model Number | CAT6KIT-A |
Catalog Number | CAT6KIT |
Lot Number | F91612 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PENUMBRA, INC. |
Manufacturer Address | ONE PENUMBRA PLACE ALAMEDA CA 94502 US 94502 |
Product Code | --- |
Date Received | 2020-03-11 |
Device Sequence No | 101 |
Device Event Key | 0 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-11 |