CENTURION SURGICAL PROCEDURE PAK 8065752200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-11 for CENTURION SURGICAL PROCEDURE PAK 8065752200 manufactured by Alcon Research, Llc - Houston.

Event Text Entries

[183075005] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[183075006] A pharmacist reported a patient experienced a corneal burn at incision edge during cataract extraction surgery. The phacoemulsification (phaco) did not work with the handpiece. The handpiece made "milk" in the patient's lens. The handpiece was removed form the eye and exchanged but the system did not work well. The system was switched to perform a combined vitrectomy/ cataract procedure due to a burned cornea and a dislocated fragment that fell into the vitreous cavity. There was corneal melting at closure of the case. The patient was hospitalized. Additional information has been requested but not received. This is one of four reports from this facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644019-2020-00137
MDR Report Key9819362
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2020-01-31
Date Mfgr Received2020-02-13
Device Manufacturer Date2019-11-14
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON RESEARCH, LLC - HOUSTON
Manufacturer Street9965 BUFFALO SPEEDWAY
Manufacturer CityHOUSTON TX 77054
Manufacturer CountryUS
Manufacturer Postal Code77054
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCENTURION SURGICAL PROCEDURE PAK
Generic NameGENERAL SURGERY TRAY (KIT)
Product CodeLRO
Date Received2020-03-11
Model NumberNA
Catalog Number8065752200
Lot Number2339243H
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LLC - HOUSTON
Manufacturer Address9965 BUFFALO SPEEDWAY HOUSTON TX 77054 US 77054


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.