3DIMENSIONS 3DM-SYS-STD

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-11 for 3DIMENSIONS 3DM-SYS-STD manufactured by Hologic, Inc.

Event Text Entries

[183261235] As of today the investigation is still in process and a follow up will be filed as needed.
Patient Sequence No: 1, Text Type: N, H10


[183261236] It was reported that when using the fingerprint scanner an electrical shock was felt by two technologists. No injury reported. A field engineer was dispatched to the site and was not able to confirm the reported issue. No voltage was measured from the fingerprint scanner. The assembly, which included the emergency stop button and fingerprint scanner was replaced. Per the customer, there has been no further reported issues since the repair.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220984-2020-00037
MDR Report Key9820256
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-11
Date of Report2020-02-17
Date Mfgr Received2020-02-17
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN FORNIERI
Manufacturer Street36 & 37 APPLE RIDGE ROAD
Manufacturer CityDANBURY, CT
Manufacturer CountryUS
Manufacturer Phone7318491
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3DIMENSIONS
Generic NameMAMMOGRAPHY SYSTEM
Product CodeOTE
Date Received2020-03-11
Model Number3DM-SYS-STD
Catalog Number3DM-SYS-STD
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC
Manufacturer Address36 & 37 APPLE RIDGE ROAD DANBURY, CT US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.