LEVEL 1 TRAUMA FAST FLOW SYSTEM H-1200 8002950

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-11 for LEVEL 1 TRAUMA FAST FLOW SYSTEM H-1200 8002950 manufactured by Smiths Medical Asd, Inc.

Event Text Entries

[183086329] Evaluation results: one level 1 trauma fast flow system was returned for investigation in fair condition. The investigator set the air regulator pressure to 5. 6 psi +0/-2psi. The h-2 pressure gauge read below the expected 270-300 mmgh. The customer reported product problem was therefore confirmed. The pressure gauges were replaced as a result. The unit passed all functional tests after this was performed. The problem source of the reported product problem was unknown. A root cause was not established.
Patient Sequence No: 1, Text Type: N, H10


[183086330] It was reported that the left chamber failed to produce an accurate reading. The product problem was observed during patient use. The product problem did not cause or contribute to any patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-01903
MDR Report Key9820805
Report SourceUSER FACILITY
Date Received2020-03-11
Date of Report2020-03-11
Date of Event2019-07-22
Date Mfgr Received2020-02-14
Device Manufacturer Date2016-06-04
Date Added to Maude2020-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1 TRAUMA FAST FLOW SYSTEM
Generic NameWARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Product CodeBSB
Date Received2020-03-11
Returned To Mfg2019-08-27
Model NumberH-1200
Catalog Number8002950
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.