MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-11 for ADVANTA 2 FRAME P1190A000016 manufactured by Hill-rom Mexico.
[186870342]
The hillrom technician found the side communication cable needed to be replaced. Per the hillrom service manual the advanta? 2 should be subject to an effective maintenance program. This will help make sure of a long, operative life for the advanta? 2 bed. The preventative maintenance will help to reduce downtime due to excessive wear failures. An annual service of the bed is advised in order to maintain its characteristics and performance. Sidecom communication system: inspect and test the communication junction box. Make sure the sidecom? Communication system features operate correctly. Inspect the communication cable, including the male and female pins in the plug. Replace as necessary. A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in december 2018. It is unknown if the facility performed any other preventative maintenance on this bed. The technician replaced the side communication cable to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10
[186870363]
Hillrom received a report from a hillrom technician stating the bed's nurse call was not functioning. The bed was located in (b)(6) at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint # (b)(4).
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3006697241-2020-00017 |
MDR Report Key | 9821789 |
Report Source | COMPANY REPRESENTATIVE,USER F |
Date Received | 2020-03-11 |
Date of Report | 2020-02-19 |
Date of Event | 2020-02-19 |
Date Mfgr Received | 2020-02-19 |
Device Manufacturer Date | 2011-01-31 |
Date Added to Maude | 2020-03-11 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LORI DAULTON |
Manufacturer Street | 1069 STATE ROUTE 46 EAST |
Manufacturer City | BATESVILLE IN 47006 |
Manufacturer Country | US |
Manufacturer Postal | 47006 |
Manufacturer Phone | 8129312280 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ADVANTA 2 FRAME |
Generic Name | BED, AC-POWERED ADJUSTABLE HOSPITAL |
Product Code | FNL |
Date Received | 2020-03-11 |
Model Number | P1190A000016 |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | HILL-ROM MEXICO |
Manufacturer Address | AVE.DEL TELEFONO NO. 200 COL. HUINALA APODACA, NUEVO LEON 66640 MX 66640 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-11 |