LEVEL 1 DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SETS DI-60HL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-12 for LEVEL 1 DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SETS DI-60HL manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[183089767] Device evaluation: returned device was received in good physical condition. During the evaluation of the device the sample was tested on a h-1200 fast flow fluid warmer and a leak was found between the tube and cap. The customer reported condition was confirmed. Problem source was traced to manufacturing. The device history record was reviewed and showed that no issues were identified that would have impacted this event.
Patient Sequence No: 1, Text Type: N, H10


[183089768] Information was received that during a pre-use check a smiths medical level 1 disposable normothermic iv administration sets was found leaking from the tube. So they refrained from using it and replaced it with another new one. No patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-01977
MDR Report Key9822079
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-12
Date of Report2020-03-11
Date of Event2019-12-01
Date Mfgr Received2020-02-10
Device Manufacturer Date2019-08-15
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1 DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SETS
Generic NameDEVICE, WARMING. BLOOD AND PLASMA, PRODUCT CODE: KZL
Product CodeKZL
Date Received2020-03-12
Returned To Mfg2019-12-20
Model NumberDI-60HL
Catalog NumberDI-60HL
Lot Number3840107
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.