TRIMA ACCEL 82460

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-12 for TRIMA ACCEL 82460 manufactured by Terumo Bct.

Event Text Entries

[186018488] Investigation: the run data file (rdf) was analyzed for this event. Per the rdf, signals indicate it is possible the platelet bag could have had an occlusion that prohibited platelets from making it out of the lrs chamber and into the collect bag. This may have overwhelmed the chamber and contributed to contamination once the occlusion was resolved at 40 minutes. Investigation is in process. A follow-up report will be provided.
Patient Sequence No: 1, Text Type: N, H10


[186018489] "the customer would like the run data file investigated to determine a possible cause for the overflow into the platelet concentrate when the additive solution was added. There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event. Donor unit #: w141619653055 the platelet collection set is not available for return because it was discarded by the customer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1722028-2020-00099
MDR Report Key9823148
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-01-07
Date Mfgr Received2020-02-19
Device Manufacturer Date2019-09-19
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSCOT HILDEN
Manufacturer Street10810 W. COLLINS AVE
Manufacturer CityLAKEWOOD CO 80215
Manufacturer CountryUS
Manufacturer Postal80215
Manufacturer Phone3032314970
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIMA ACCEL
Generic NameTRIMA PLT SMPLR AUTOPAS, MULTI PLS, RBC
Product CodeGKT
Date Received2020-03-12
Catalog Number82460
Lot Number1909171230
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTERUMO BCT
Manufacturer AddressLAKEWOOD CO 80215 US 80215


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-12