INVACARE PERFECTO2 V OXYGEN CONCENTRATOR IRC5PO2V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2020-03-12 for INVACARE PERFECTO2 V OXYGEN CONCENTRATOR IRC5PO2V manufactured by Invacare Florida.

Event Text Entries

[183731050] The dealer advised that the customer is not a smoker, and it didn't appear that the unit had been tampered with. Photographs of the concentrator were provided, which show deformation to the left sieve bed cap and discoloration of the attached tubing, indicating that an overheating event had occurred. Additionally, the photos show discoloration and deformation of the cabinet's exterior near the inlet filter and cabinet filter door, indicating that the overheating event exited the shroud. Based on this evidence that the overheating event exited the shroud, this incident is being reported to the fda out of an abundance of caution. This failure mode has been previously identified and investigated. The investigation activities concluded that the failure is unlikely to occur when the system is operating under normal conditions or a single fault condition. Rather, the system likely has to experience multiple faults to result in the failure observed. Even though the failure is unlikely to occur, components of the irc5po2v concentrator have been updated to prevent potential recurrence of this issue. The subject concentrator was manufactured prior to implementation of these updates. The device has been returned to invacare for further evaluation. Once the evaluation results are available, a supplemental record will be filed.
Patient Sequence No: 1, Text Type: N, H10


[183731051] The dealer reported that the irc5po2v concentrator had signs that a fire had occurred, though a fire was not actually witnessed. The dealer stated that the inside of the unit had melted in the area of the left sieve bed and pe valve.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1031452-2020-00006
MDR Report Key9823634
Report SourceOTHER
Date Received2020-03-12
Date of Report2020-02-18
Date Mfgr Received2020-02-18
Device Manufacturer Date2016-10-10
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44035
Manufacturer CountryUS
Manufacturer Postal44035
Manufacturer Phone8003336900
Manufacturer G1INVACARE FLORIDA
Manufacturer Street2101 EAST LAKE MARY BLVD
Manufacturer CitySANFORD FL 32773
Manufacturer CountryUS
Manufacturer Postal Code32773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINVACARE PERFECTO2 V OXYGEN CONCENTRATOR
Generic NameGENERATOR, OXYGEN, PORTABLE
Product CodeCAW
Date Received2020-03-12
Returned To Mfg2020-03-04
Model NumberIRC5PO2V
Catalog NumberIRC5PO2V
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerINVACARE FLORIDA
Manufacturer Address2101 EAST LAKE MARY BLVD SANFORD FL 32773 US 32773


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.