3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-12 for 3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355 manufactured by 3m Health Care.

Event Text Entries

[187708063] Patient information: no information was provided. The sample was evaluated. Sample evaluation showed a leak in the drip chamber due to a solvent bond failure. Corrective action was issued to address solvent bond leaks. Lot hx8341 was produced post corrective action implementation. 3m will continue to monitor.
Patient Sequence No: 1, Text Type: N, H10


[187708064] A hospital employee reported that a 3m? Ranger? High flow disposable set (model 24355) was primed with normal saline and loaded into a ranger? Rapid pressure infuser during an mtp procedure on (b)(6) 2020. The employee alleged that upon initiation of the pressure infuser, blood started to leak around the filter chamber and over the floor. The flow rate at the time was high. The employee reported that the active bleeding patient needed a massive transfusion and care was delayed. The tubing reportedly was switched, a new set was primed, and the mtp was restarted. No harmful exposure to blood was specified. No patient or staff injury was specified. There were reportedly no issues with the disposable set during priming. The patient's current condition is unknown. The ranger? Pressure infusor was sent to the facility's biomed for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00015
MDR Report Key9823903
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-02-14
Date Mfgr Received2020-02-14
Device Manufacturer Date2019-03-05
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Manufacturer G1NYPRO PRECISION ASSEMBLIES
Manufacturer StreetSOR JUANA INES DE LA CRUZ CD. INDUSTRIAL
Manufacturer CityTIJUANA, 22444
Manufacturer CountryMX
Manufacturer Postal Code22444
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? RANGER? HIGH FLOW DISPOSABLE SET
Generic NameHIGH FLOW DISPOSABLE TUBING
Product CodeBSB
Date Received2020-03-12
Returned To Mfg2020-02-24
Model NumberN/A
Catalog Number24355
Lot NumberHX8341
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.