TECNIS 1 MULTIFOCAL ZLB00 ZLB00U0230

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-12 for TECNIS 1 MULTIFOCAL ZLB00 ZLB00U0230 manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[183123580] Date of event: unknown, not provided; best estimate is between (b)(6) 2019 and (b)(6) 2020. (b)(4). Device evaluation: the product was received in a lens case for an alcon tfat00 (serial number: (b)(4), and a diopter of 23. 0d). Visual inspection under magnification revealed no cosmetic defects. The complaint issue could not be confirmed, and no product deficiency was identified. Manufacturing records review: the manufacturing records for the product were reviewed. The product was manufactured and released according to specification. Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number. Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[183123581] It was reported that a zlb00 intraocular lens (iol) was implanted in the patient's eye on (b)(6) 2019. It was later explanted on (b)(6) 2020 due to mechanical failure. Incision enlargement along with suture was used. There was no complication or patient injury reported. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9614546-2020-00119
MDR Report Key9823952
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-12
Date of Report2020-03-12
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-17
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer StreetVAN SWIETENLAAN 5
Manufacturer CityGRONINGEN 9728NX
Manufacturer CountryNL
Manufacturer Postal Code9728 NX
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameTECNIS 1 MULTIFOCAL
Generic NameMULTIFOCAL IOLS
Product CodeMFK
Date Received2020-03-12
Returned To Mfg2020-03-05
Model NumberZLB00
Catalog NumberZLB00U0230
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E ST ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-12
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.