DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MT20649-3 STR-JP-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2020-03-12 for DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MT20649-3 STR-JP-001 manufactured by Dexcom, Inc..

Event Text Entries

[183131881] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[183131882] It was reported that error icon displayed occurred. No product or data was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient Sequence No: 1, Text Type: D, B5


[188146965] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188146966] It was reported that error icon displayed occurred. The product was evaluated. An external visual inspection was performed and passed. Charge and boot test was performed and passed. The log was downloaded and reviewed finding no errors related to the complaint. The allegation was not confirmed. The probable cause could not be determined. No injury or medical intervention was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004753838-2020-024300
MDR Report Key9824449
Report SourceDISTRIBUTOR,FOREIGN
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-02-12
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS ASHLEY SPOTO
Manufacturer Street6340 SEQUENCE DR.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582000200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Generic NameCONTINUOUS GLUCOSE MONITOR
Product CodeMDS
Date Received2020-03-12
Model NumberMT20649-3
Catalog NumberSTR-JP-001
Lot NumberNI
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDEXCOM, INC.
Manufacturer Address6340 SEQUENCE DR. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12