[183638853] Vascular pta balloon rupture. Max inflation pressure not reached. Provider wasn't able to suck the balloon down due to rupture. Upon removal the tip of the balloon broke off, f/u ct scan showed the fragment in the femoral mein. Balloon burst even though rated burst pressure hadn't been achieved. Upon withdrawal of the balloon through the sheath in the femoral vein, the distal part of the balloon separated and lodged in the vein. We believe this happened because the unwrapped balloon material caught on the sheath.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5093705
MDR Report Key9824485
Date Received2020-03-11
Date of Report2020-02-28
Date of Event2020-02-19
Date Added to Maude2020-03-12
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NamePTA BALLOON DILITATION CATHETER CHARGER 12MM X 60MM 75CM SHAFT
Generic NameCATHETER, ANGIOPLASTY, PERIPHERAL TRANSLUMINAL
Product CodeNVM
Date Received2020-03-11
Model NumberH74939206120670
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION

Device Sequence Number: 101

Brand NameCHARGER 12MM X 60MM 75 SHAFT
Product Code---
Date Received2020-03-11
Device Sequence No101
Device Event Key0
ManufacturerBOSTON SCIENTIFIC