TRANSFORM SUPER COMPLIANT 4MM X 10MM- US M003SSC04100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-12 for TRANSFORM SUPER COMPLIANT 4MM X 10MM- US M003SSC04100 manufactured by Stryker Neurovascular Cork.

Event Text Entries

[183295716] It was reported that during the procedure, the balloon catheter(subject device) failed to inflate. There was a hole in the distal segment of the balloon. There were no reported clinical consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008881809-2020-00063
MDR Report Key9824725
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-02-20
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TARA LOPEZ
Manufacturer Street47900 BAYSIDE PARKWAY
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal94538
Manufacturer Phone5104132500
Manufacturer G1STRYKER NEUROVASCULAR CORK
Manufacturer StreetIDA INDUSTRIAL ESTATE MODEL FARM ROAD
Manufacturer CityCORK NA
Manufacturer CountryIE
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSFORM SUPER COMPLIANT 4MM X 10MM- US
Generic NameCATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Product CodeMJN
Date Received2020-03-12
Returned To Mfg2020-03-04
Model NumberM003SSC04100
Catalog NumberM003SSC04100
Lot Number21667486
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER NEUROVASCULAR CORK
Manufacturer AddressIDA INDUSTRIAL ESTATE MODEL FARM ROAD CORK NA IE NA


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12