ZOLL IVTM THERMOGARD XP 8700-0650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-12 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.

Event Text Entries

[188240110] The reported complaint of the thermogard ivtm system (serial (b)(4)) displaying error message "tcm id:05" (coolant diff failure) was confirmed during initial functional test and event log review. The investigation findings revealed a dirty and corroded sensors of lm-34 probe. The root cause of the reported error was due to an improper maintenance. Unrelated to the reported complaint, a rusted cold well and worn white rollers, missing cold well cap, a weee label, a pan drip and rear bracket screws were observed on the returned thermogard ivtm system during visual inspection. These types of rusted, worn and missing components are likely due to improper maintenance and/or due to normal wear and tear. Cold well, white rollers, cold well cap, a weee label, a pan drip and rear bracket screws were replaced. The thermogard ivtm system was manufactured in march 2011 and it is 9 years old, well past beyond its expected serviceable life of 5 years during archive review, multiple tcm id:05 error messages were observed on the reported event date. Thus, confirming the reported complaint. Initial functional testing on the themogard system failed due to error message "tcm id:05" on the display. To remedy "tcm id:05", the lm34 temperature sensor assembly needs to be replaced. Upon service completion, the thermogard xp ivtm system will be further tested to full specification. Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for thermogard ivtm system with serial number (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188240111] Customer reported that the thermogard xp ivtm system (serial #(b)(4)) displayed "tcm id:05" (coolant diff failure) error message upon powering on. No consequences or impact to patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00244
MDR Report Key9824888
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-02-18
Date Mfgr Received2020-02-18
Device Manufacturer Date2011-03-24
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM THERMOGARD XP
Generic NameTHERMAL REGULATING SYSTEM
Product CodeNCX
Date Received2020-03-12
Returned To Mfg2020-02-27
Model Number8700-0650
Catalog Number8700-0650
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.