3M MULTIPORE DRY SURGICAL TAPE N/A 3730-0

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other,use report with the FDA on 2020-03-12 for 3M MULTIPORE DRY SURGICAL TAPE N/A 3730-0 manufactured by 3m Health Care.

Event Text Entries

[184050746] Patient identifier, sex, ethnicity, race: information was not provided by reporter. Sample has not yet been returned for evaluation. Lot number was not provided. The tape was used to secure an endotracheal tube (brand unspecified) while the patient was at the neonatal station in an incubator at 30. 5 degrees c and 45% humidity. The tape was used for approximately 24 hours prior to observation of the issue, and was reported to have stretched and lost adherence; adhesive was observed on the skin.
Patient Sequence No: 1, Text Type: N, H10


[184050747] Received from (b)(6): customer reported 3m? Multipore? Dry surgical tape that was used to secure an endotracheal tube for a (b)(6) old infant who was in an incubator in a neonatal station slipped, stretched and lost adherence approximately 24 hours after application. Adhesive was observed on the patient's skin. The tube was dislocated and re-taped several times before it was re-affixed with a non-3m brand tape. No extubation or patient injury occurred.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00016
MDR Report Key9825410
Report SourceHEALTH PROFESSIONAL,OTHER,USE
Date Received2020-03-12
Date of Report2020-03-11
Date of Event2020-02-10
Date Mfgr Received2020-02-14
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Manufacturer G13M JAPAN PRODUCT LIMITED
Manufacturer Street5500 OOAZA-OSANAGI HIGASHINE-SHI
Manufacturer CityYAMAGATA, 999-3737
Manufacturer CountryJA
Manufacturer Postal Code999-3737
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M MULTIPORE DRY SURGICAL TAPE
Generic Name3M MULTIPORE DRY TAPE
Product CodeKGX
Date Received2020-03-12
Model NumberN/A
Catalog Number3730-0
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-12