TRIMA ACCEL 82406

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-12 for TRIMA ACCEL 82406 manufactured by Terumo Bct.

Event Text Entries

[186017896] Investigation: the run data file was analyzed for the event. The analysis of the run data file did not find a conclusive cause for the higher-than- expected wbc content in the platelet product reported for this collection. No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended. Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related. Investigation is in process. A follow up report will be provided.
Patient Sequence No: 1, Text Type: N, H10


[186017897] The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product. There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event. Donor unit #: w04702003297900g the platelet collection set is not available for return because it was discarded by the customer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1722028-2020-00105
MDR Report Key9827638
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-12
Date of Report2020-03-12
Date of Event2020-02-18
Date Mfgr Received2020-02-19
Device Manufacturer Date2019-12-04
Date Added to Maude2020-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSCOT HILDEN
Manufacturer Street10810 W.COLLINS AVE
Manufacturer CityLAKEWOOD CO 80215
Manufacturer CountryUS
Manufacturer Postal80215
Manufacturer Phone3032314970
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIMA ACCEL
Generic NameTRIMA ACCEL LRS PLATELET, PLASMA, RBC SET
Product CodeGKT
Date Received2020-03-12
Model Number82406
Catalog Number82406
Lot Number1912063130
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTERUMO BCT
Manufacturer AddressLAKEWOOD CO 80215 US 80215


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.