ULTRAXX NEPHROSTOMY BALLOON AND SET UNBS-10-15-CS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-13 for ULTRAXX NEPHROSTOMY BALLOON AND SET UNBS-10-15-CS manufactured by Cook Inc.

Event Text Entries

[183355054] Occupation: other non-healthcare professional: buyer. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[183355055] It was reported, during a percutaneous nephrolithotomy (pcnl) using a ultraxx nephrostomy balloon and set, the stopcock was letting out pressure and fluid. When starting to inflate the balloon, the stopcock was letting out pressure and fluids (1/2 contrast and 1/2 saline). The stopcock was leaking where it tightens to the catheter. The balloon was inflated/deflated two times. The stopcock was then removed and the inflation device was connected directly to the balloon without the stopcock. The procedure was successfully completed. According to the initial reporter, the patient did not experience any adverse effects due to the alleged malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00607
MDR Report Key9831048
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-13
Date of Report2020-04-03
Date of Event2020-03-10
Date Mfgr Received2020-03-25
Device Manufacturer Date2020-01-06
Date Added to Maude2020-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameULTRAXX NEPHROSTOMY BALLOON AND SET
Generic NameLJE CATHETER, NEPHROSTOMY
Product CodeLJE
Date Received2020-03-13
Returned To Mfg2020-03-19
Catalog NumberUNBS-10-15-CS
Lot Number10246282
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-13

© 2026 FDA.report