HEATER-COOLER SYSTEM 3T 16-02-85

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-13 for HEATER-COOLER SYSTEM 3T 16-02-85 manufactured by Livanova Deutschland Gmbh.

Event Text Entries

[183750101] Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices. The z number is z-2076/2081-2015. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Through further follow-up communication with the chief perfusionist livanova learned that the water in the heater-cooler systems 3t in use is changed every day and they are stored dry. This is not in alignment with current instruction for use however reportedly the devices are very clean and there is no sign of biofilm. The devices are located inside the operating theatre during use. We are currently waiting to know the result of microbial sampling performed at customer site. Reportedly the tests are still in progress. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[183750103] A male patient born on (b)(6) undergone a cardiac surgery on (b)(6) 2019. Patient resulted to be infected with m. Avium intracellulare complex at the sternal tissue. This patient infection was previously reported under importer (b)(4) and manufacturer mw-07589. Following to additional information received from the hospital a patients mapping has been done and separated cases have been created.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611109-2020-00186
MDR Report Key9831309
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-13
Date of Report2020-03-13
Date of Event2019-06-12
Date Mfgr Received2020-02-13
Device Manufacturer Date2017-05-02
Date Added to Maude2020-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactENRICO GRECO
Manufacturer Street14401 W. 65TH WAY
Manufacturer CityARVADA CO 80004
Manufacturer CountryUS
Manufacturer Postal80004
Manufacturer G1LIVANOVA DEUTSCHLAND GMBH
Manufacturer StreetLINDBERGHSTR. 25
Manufacturer CityMUNICH 80309
Manufacturer CountryGM
Manufacturer Postal Code80309
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-2076/2081-2015
Event Type3
Type of Report3

Device Details

Brand NameHEATER-COOLER SYSTEM 3T
Generic NameCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Product CodeDWC
Date Received2020-03-13
Model Number16-02-85
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerLIVANOVA DEUTSCHLAND GMBH
Manufacturer AddressLINDBERGHSTR. 25 MUNICH 80309 GM 80309


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.