PENUMBRA SYSTEM ASPIRATION PUMP MAX 220 PMX220

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2020-03-13 for PENUMBRA SYSTEM ASPIRATION PUMP MAX 220 PMX220 manufactured by Penumbra, Inc..

Event Text Entries

[188418499] This device is available for return. A follow up mdr will be submitted upon completion of the device investigation.
Patient Sequence No: 1, Text Type: N, H10


[188418503] The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 220 (pump max) and a penumbra system ace 68 reperfusion catheters (ace68). During the procedure, the pump max aspiration pressure would not increase beyond -20 inhg resulting in the decrease of suction force for the thrombus; therefore, the pump max was removed. It was reported that multiple attempts were made to troubleshoot the pump max but were unsuccessful. The procedure was completed using manual aspiration with a 60cc syringe. There was no report of an adverse effect to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005168196-2020-00362
MDR Report Key9832356
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2020-03-13
Date of Report2020-02-21
Date of Event2020-01-20
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. VERONICA FARRIS
Manufacturer StreetONE PENUMBRA PLACE
Manufacturer CityALAMEDA CA 94502
Manufacturer CountryUS
Manufacturer Postal94502
Manufacturer Phone5107483200
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NamePENUMBRA SYSTEM ASPIRATION PUMP MAX 220
Generic NameNRY
Product CodeNRY
Date Received2020-03-13
Model NumberPMX220
Catalog NumberPMX220
Lot NumberF42669-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerPENUMBRA, INC.
Manufacturer AddressONE PENUMBRA PLACE ALAMEDA CA 94502 US 94502

Device Sequence Number: 101

Product Code---
Date Received2020-03-13
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-13

© 2024 FDA.report